Many sensors produce an output in terms of units other than volts – in these cases, an additional scaling factor must be applied to the scale calculation. This scaling factor can be calculated once some basic information about the sensor is available. In this example we will calculate the ANALOG:SCALE parameter for a load cell used to measure tension and compression so that we record the output in the same units that are used to calibrate the load cell. For this example we will use a Sensotec, Inc., model 31 Load Cell rated at 50lbs. The load cell performance sheet provides the following information for this device:
|
Output |
2mV/V |
|
Excitation |
10.0 VDC |
The load cell output is specified in terms of volts output per volt of excitation at full load. In this case, the manufacturer specifies a 10.0 Volt excitation voltage, so the load cell output will be 20mVat full load, which, for this load cell, is 50 pounds. We now have enough information to calculate the sensor calibration factor:
Many modern ADC sampling devices can be programmed to use different input ranges by changing the ADC gain. If we use an ADC_gain of x8 in the above scale calculations, we can improve the measurement resolution to about 1.5 lbs. This resolution can be further improved by adding an additional gain stage in between the load cell and the ADC.